The A/T 2000 XR fully automatic film processor develops all sizes; panoramic, cephalometric, TMJ, occlusal and periapicals. Processing time, chemistry temperature and chemistry replenishment are automatically monitored and adjusted. Archival quality films are processed in 5 ½ minutes. Plus, there is an Endo Speed for quick-look films in only 2 ½ minutes.
Electrical: 115 Volt, 60 Hz, 8 Amps. Use a 3 prong, grounded outlet. A separate dedicated 15 Amp line is recommended.
Water Flow: 1/2 gallon per minute (while processing film). Water softeners should not be used.
Water Pressure: 80 psi maximum/30 psi minimum. Water source must have a manual shut-off near processor.
Water Connections: Adapters are provided to connect to 1/2" MPT or 3/4" male garden hose fitting. The supplied tubing at rear of the processor is 6' long.
Water Backflow Prevention: A vacuum breaker is not required because the processor has a 1" air gap between water supply inlet and top of operating water level in wash tank.
Water Temp: 55°F (13°C) to 80°F (27°C). A mixing valve is not required if water temperature is within this range.
Drain: 1 ½ diameter PVC standpipe, open at top, connected to 1 ½" PVC trap. DO NOT DRAIN INTO COPPER OR BRASS. The top of the standpipe must be 6" below the bottom of the countertop.
Dimensions (HxWxD): 18" x 15" x 25" with leveling base.
Weight: 90 lbs. with water and chemistry (75 lbs. empty).
Ventilation: Room air temperature must not exceed 80°F (27°C) during processing. Because heat is generated during processing, adequate ventilation is necessary to maintain darkroom temperature.
|